13.12 Temperature variations
Temperature variations can significantly influence the measuring results. To reduce this
influence, it is recommended to check and store the probe constant regularly.
13.13 Cleaning
The plastic parts, the display unit as well as the painted parts of the instrument should be cleaned using a
slightly wet (watered) fabric. Clean the air cushion pads using a clean fabric, slightly soaked with alcohol.
14 Technical specifications
Vectra-Touch
300
Vectra-Touch
600
Vectra-Touch
1000
Measuring range mm 305 610 1016
Application range mm 567 872 1278
Max. resolution mm 0.001
Max. permissible error µm
2.5 + L(mm)/400
Repeatability µm
1 (diameters: 2)
Max. manual displacement speed mm/s 1000
Max. motorized displacement speed mm/s 150
Measuring force N 0.5 .. 1.8
Power supply Incorporated battery pack, rechargeable
Autonomy h > 8
Overall squareness deviation using probe µm
4 6 10
Data output 2 x RS232 C and 2 x USB (A and B)
Operational temperature °C +10 .. +40
Storage temperature °C -10 .. +40
Total height mm 645 950 1357
Weight kg 16 19 23
The values of max. permissible error and repeatability are valid only when using the standard probe (measuring insert with tungsten carbide
ball, 4 mm, L = 90 mm) at temperature of 20 ± 0.5 °C and relative humidity of 50 ±5%.
Mestra -Touch
300
Mestra -Touch
600
Mestra -Touch
1000
Measuring range mm 305 610 1016
Application range mm 567 872 1278
Max. resolution mm 0.0001
Max. permissible error µm
1.4 + L(mm)/400
Repeatability µm
0.5 (diameters: 1)
Max. manual displacement speed mm/s 1000
Max. motorized displacement speed mm/s 150
Measuring force N 0.5 .. 1.8
Power supply Incorporated battery pack, rechargeable
Autonomy h > 8
Overall squareness deviation using probe µm
4 6 10
Data output 2 x RS232 C and 2 x USB (A and B)
Operational temperature °C +10 .. +40
Storage temperature °C -10 .. +40
Total height mm 645 950 1357
Weight kg 16 19 23
The values of max. permissible error and repeatability are valid only when using the standard probe (measuring insert with tungsten carbide
ball, 4 mm, L = 90 mm) at temperature of 20 ± 0.5 °C and relative humidity of 50 ±5%.